125 Pocket Calibrator
Model 125 Pocket Calibrator is a portable, lightweight
simulator designed to supply millivolt-per-volt level signals
for testing, calibrating, and troubleshooting load cell/scale
indicators. Precise output references for 0, 1, 2, or 3 mV/V
are achieved by using a metal film resistor network,
discrete wire wound resistors, and a 2-pole, 4-position
rotary switch.
The 350 Ω input and output impedance matches typical
strain gage devices. Four permanent binding posts,
integral to the rugged palm-size case, provide connection
points for the indicator or transmitter.
Model 125 units substitute for platform or scale
transducers. Lightweight construction, compact size,
and good accuracy make the Model 125 Calibrator an
excellent choice for calibrating, spot-checking, or trouble
shooting any electronic weigh system
- ◉ Low cost, on-site calibration and servicing
- ◉ 4 selectable ranges: 0,1, 2, and 3 mV/V
- ◉ Calibrate any strain gage based transducer indicator
- ◉ Rugged, pocket size case
- Portable load cell/weigh system simulator
PARAMETER |
VALUE |
Output Accuracy |
0.02% of selected range |
Accuracy Stability |
better than 0.01% in 24 hours better than 0.02% in 1 year |
Zero Stability |
less than 3 μV |
Span TC |
±10 ppm/°C |
Input Impedance (Excitation) |
1000 Ω ±0.05% |
Output Impedance (Signal) |
350 Ω ±0.08% |
Output Ranges |
4 steps: 0, 1, 2, and 3 mV/V |
Input Voltage Level |
25 VDC max |
Operating Temperature Range |
32°F to 120°F (0°C to 50°C) |
Dimensions, LxWxH |
3.3 × 2.35 × 1.4 in. |
Unit Weight |
4.8 oz |